کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1217078 1494177 2007 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A simplified method for the measurement of urinary free cortisol using LC–MS/MS
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
A simplified method for the measurement of urinary free cortisol using LC–MS/MS
چکیده انگلیسی

The measurement of 24 h urinary free cortisol is used in the investigation of patients with symptoms of hypercortisolism. Many different methods have been published for the measurement of cortisol, but most of these methods involve cumbersome pre-extraction of the cortisol prior to analysis. We have developed a method using in-well protein precipitation which serves to clean up the sample without requiring lengthy sample preparation. A Shimadzu SIL-HT autosampler was used to inject 50 μL of extract onto a Phenomemex® Gemini C18 guard column attached to a Waters™ Xbridge C18 column. The eluant was introduced directly into a Waters™ Quattro Micro tandem mass spectrometer. The method was found to be linear up to 3448 nmol/L with a lower limit of detection of 5.3 nmol/L. Precision and accuracy were acceptable, and no interference was noted from compounds such as prednisolone or fenofibrate. This assay was compared to a previously published method, which uses solid phase extraction prior to LC–MS/MS analysis. We have developed a simplified, robust assay for the quantitation of urinary free cortisol that will increase the throughput of the assay and avoid the use of neurotoxic solvents such as dichloromethane.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography B - Volume 858, Issues 1–2, 15 October 2007, Pages 27–31
نویسندگان
, , , ,