Modified high-performance liquid chromatography with electrochemical detection method for plasma measurement of levodopa, 3-O-methyldopa, dopamine, carbidopa and 3,4-dihydroxyphenyl acetic acid
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Modified high-performance liquid chromatography with electrochemical detection method for plasma measurement of levodopa, 3-O-methyldopa, dopamine, carbidopa and 3,4-dihydroxyphenyl acetic acid
چکیده انگلیسی

Plasma measurements of levodopa and its major metabolites including dopamine and 3-O-methyldopa have been limited by cumbersome methods and poor sensitivity within relatively narrow ranges of plasma levels. We now report a modification of an HPLC method that permits concomitant measurements of a wide range of concentrations of levodopa, dopamine (DA), carbidopa, 3-O-methyldopa (3-OMD) and 3,4-dihydroxyphenyl acetic acid (DOPAC) from one HPLC injection. The recoveries ranged from 77 to 107% with an intra-day precision around 5% (CV) and inter-day CV's about 10–20%. This validated method will simplify pharmacokinetic studies of levodopa and its metabolites for mechanistic studies or therapeutic clinical monitoring which play a crucial role in development of strategies to prolong motor benefits from individual doses and reduce involuntary movements called dykinesias.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography B - Volume 836, Issues 1–2, 19 May 2006, Pages 120–123