• A novel CL method for bisphenol A using gold nanoparticle–Co2+–HCO4− is developed.
• The present CL system exhibited high selectivity and sensitivity toward bisphenol A.
• The proposed method has been validated by determining bisphenol A in real samples.
In this work, we utilized the nonionic fluorosurfactant-capped gold nanoparticles (GNPs) as a novel chemiluminescence (CL) probe for the determination of trace bisphenol A. Bisphenol A can induce a sharp decrease in CL intensity from the GNP–Co2+–peroxymonocarbonate (HCO4−) system. Under the selected experimental conditions, a linear relationship was obtained between the CL intensity and the logarithm of concentration of bisphenol A in the range of 0.05–50 μM (R2 = 0.9936), and the detection limit at a signal-to-noise ratio of 3 for bisphenol A was 10 nM. The applicability of the proposed method has been validated by determining bisphenol A in real polycarbonate samples with satisfactory results. The recoveries for bisphenol A in spiked samples were found to be between 94.4% and 105.0%. The relative standard deviation (RSD) for 12 repeated measurements of 0.5 μM bisphenol A was 2.2%. The proposed method described herein was simple, selective and obviated the need of extensive sample pretreatment.
Figure optionsDownload as PowerPoint slide
Journal: Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy - Volume 128, 15 July 2014, Pages 393–397