کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1569563 1514259 2007 9 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Monte Carlo simulations of stray neutron radiation exposures in proton therapy
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه مهندسی انرژی انرژی هسته ای و مهندسی
پیش نمایش صفحه اول مقاله
Monte Carlo simulations of stray neutron radiation exposures in proton therapy
چکیده انگلیسی

Proton radiotherapy patients receive unwanted stray neutron radiation generated in the treatment apparatus. The literature on this neutron radiation is sparse and disparate and pertains mainly to unique or antiquated systems. The present work reports on neutron dose equivalent per therapeutic absorbed dose (H/D) in a contemporary proton therapy system. We applied the Monte Carlo method to predict neutron exposure as a function of treatment field size and location in the treatment room. For a 250-MeV proton beam, H/D values at isocenter increased from 13 to 20 mSv Gy−1 as the collimating aperture was decreased from 18 × 18 to 0 × 0 cm2. H/D values generally decreased with distance from isocenter (20 mSv Gy−1 at isocenter versus 0.9 mSv Gy−1 at a distance of 150 cm). Neutron spectral fluence calculations revealed that >50% of the dose equivalent at all locations considered was from neutrons >10 MeV.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Nuclear Materials - Volume 361, Issues 2–3, 15 April 2007, Pages 289–297
نویسندگان
, , , , ,