کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1648516 1007559 2010 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
New material for orthopedic implants: Electrochemical study of nickel free P558 stainless steel in minimum essential medium
موضوعات مرتبط
مهندسی و علوم پایه مهندسی مواد فناوری نانو (نانو تکنولوژی)
پیش نمایش صفحه اول مقاله
New material for orthopedic implants: Electrochemical study of nickel free P558 stainless steel in minimum essential medium
چکیده انگلیسی

Nickel, a component of stainless steels (SS) applied in orthopedic implants may cause allergic processes in human tissues. P558 nickel free SS was studied to verify its viability as a substitute for stainless steel containing nickel. Its performance is compared to ISO 5832-9 and F138 most used nowadays grades in implants fabrications, in minimum essential medium, MEM, at 37 °C. Potentiodynamic polarization curves, electrochemical impedance spectroscopy (EIS), scanning electron microscopy (SEM) and “in vitro” cytotoxicity were used as techniques. From the electrochemical point of view P558 SS is comparable to ISO 5832-9 SS in MEM. It remains passivated until the transpassivation potential, above which generalized corrosion occurs. F138 presents pitting corrosion at 370 mV/SCE. The cytotoxicity results showed that P558, ISO 5832-9 and F138 do not present cytotoxic character. Therefore, these results suggest that P558 SS can be applied in orthopedic implants.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Materials Letters - Volume 64, Issue 13, 15 July 2010, Pages 1476–1479
نویسندگان
, , , , ,