• A fiducial marker system was used to measure slice geometry in 3D FIBSEM processes.
• The slice thickness variation was non-systematic and up to ±50% target thickness.
• It is likely that both instrument and material instabilities causes this variation.
• Using assumed thickness in reconstruction causes large uncertainty in measurements.
Three-dimensional reconstructions of microstructures produced by focused ion beam (FIB) milling usually assume a uniform slice thickness with flat and parallel surfaces. Measurement of the actual slice thickness and profile is difficult, and is often simply ignored. This paper reports the use of artificial 3D structures of known geometry to enable the full 3D profile of a sequence of slices produced by FIB to be measured for the first time. A transient period at the beginning of a milling process is observed in which the actual slice thickness varies by as much as ±50% from the target thickness (with significantly greater error near the base of the slice), before settling to a ±20% variation as the milling progresses. Although SEM images appear to show flat milled surfaces perpendicular to the top surface, the development of a curved, tapering milled surface is also observed. This profile is then maintained through the milling process with the bottom of the slice lagging the top by up to three slice thicknesses.
Journal: Ultramicroscopy - Volume 139, April 2014, Pages 20–28