کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1875933 1532093 2015 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A plastic scintillation counter prototype
ترجمه فارسی عنوان
نمونه اولیه ضد انفجاری پلاستیکی
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
چکیده انگلیسی


• A new device of plastic scintillation counter was developed to measure beta emitters.
• High sensitivity with low detection limit was performed for a tritium compound.
• A detection limit of tritium was 0.01 Bq mL−1 for a 10 h measurement.
• A plastic scintillation counter generated no radioactive organic waste fluid.
• A plastic scintillation counter could analyze qualitatively and quantitatively.

A new prototype device for beta-ray measurement, a plastic scintillation counter, was assembled as an alternative device to liquid scintillation counters. This device uses plastic scintillation sheets (PS sheets) as a sample applicator without the use of a liquid scintillator. The performance was evaluated using tritium labeled compounds, and good linearity was observed between the activity and net count rate. The calculated detection limit of the device was 0.01 Bq mL−1 after 10 h measurement for 2 mL sample.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 104, October 2015, Pages 175–180
نویسندگان
, ,