Charge-coupled device (CCD) digital cameras have been used to expand the application area of conventional dosimeters to two-dimensional dose measurements. Although film dosimeters are normally read with scanners, CCD reading has several advantages over scanner reading: quick reading, interference filters, and no influence from the film orientation at the reading. Moreover, the lens and light source are changeable, which makes it possible to use a wide selection of dose ranges and spatial resolutions from microscopic to macroscopic.
► CCD cameras were used for GafChromic film reading.
► The advantage over scanner was quick reading.
► Quick reading made multiple exposures easy, which decreased the fluctuation.
► A band pass filter increased the sensitivity of EBT2 higher than red LED scanner.
► No influence from the EBT2 film orientation was confirmed
Journal: Applied Radiation and Isotopes - Volume 70, Issue 7, July 2012, Pages 1214–1218