Application of the Monte Carlo method for the calibration of an in situ gamma spectrometer
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Application of the Monte Carlo method for the calibration of an in situ gamma spectrometer
چکیده انگلیسی

A MCNP model was developed for the efficiency calibration of an in situ gamma ray spectrometry system based on a high purity germanium (HPGe) detector. The detector active crystal volume was adjusted semi-empirically against experimental measurements. Calculated full energy peak efficiency curves, over the photon energy range between 50 keV and 5 MeV, are presented for surface and slab source configurations. The effect of different collimator apertures and the contribution of different HPGe crystal regions in the detector response are also shown.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 68, Issues 7–8, July–August 2010, Pages 1441–1444