کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1878233 1042289 2009 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Syringe shape and positioning relative to efficiency volume inside dose calibrators and its role in nuclear medicine quality assurance programs
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Syringe shape and positioning relative to efficiency volume inside dose calibrators and its role in nuclear medicine quality assurance programs
چکیده انگلیسی

A careful analysis of geometry and source positioning influence in the activity measurement outcome of a nuclear medicine dose calibrator is presented for 99mTc. The implementation of a quasi-point source apparent activity curve measurement is proposed for an accurate correction of the activity inside several syringes, and compared with a theoretical geometric efficiency model. Additionally, new geometrical parameters are proposed to test and verify the correct positioning of the syringes as part of acceptance testing and quality control procedures.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 67, Issue 6, June 2009, Pages 1104–1109
نویسندگان
, , , ,