کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1878659 1532110 2014 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Evaluation of three electrodeposition procedures for uranium, plutonium and americium
ترجمه فارسی عنوان
ارزیابی سه روش الکترواستاتیک برای اورانیوم، پلوتونیوم و آمریکیم
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
چکیده انگلیسی


• Four different electrodeposition procedures were performed and compared.
• Temperature changes were monitored with time and different currents.
• pH changes were monitored with time and different currents.
• Percentage deposition was monitored with time and different currents.

While both mass spectrometry and alpha-particle spectrometry have been widely used for measuring alpha activities, the former is preferred since many laboratories are not equipped with ICP-MS and MC-ICP-MS systems. In this study, three electrodeposition techniques using ammonium chloride, ammonium oxalate and ammonium sulphate as electrolyte solutions were applied and evaluated for the preparation of uranium, plutonium and americium sources for alpha-particle spectrometry. Changes in pH and temperature throughout the electrodeposition process were measured every 15 min, together with percentage deposition every 30 min. The percentage deposition in each method was checked at 300, 400 and 500 mA, and the optimised time and current were determined.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Applied Radiation and Isotopes - Volume 87, May 2014, Pages 233–237
نویسندگان
, , , ,