A methodology for choosing parameters for ESR readout of alanine dosimeters for radiotherapy
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
A methodology for choosing parameters for ESR readout of alanine dosimeters for radiotherapy
چکیده انگلیسی

Spectrometer settings for ESR readout of alanine dosimeters for radiotherapy have been investigated. Several ESR parameters were studied and determined. The main reason for this work is to choose the suitable parameters to increase signal-to-noise ratio and to reduce the uncertainty on ESR readout, which is one of the main components of uncertainty of alanine/ESR dosimetry system for radiotherapy. The new spectrometer settings have been applied for ESR readout of alanine dosimeters irradiated from 1 to 10 Gy. A higher signal-to-noise ratio has been achieved compared to our old spectrometer settings. The extended uncertainty (k=2) has been evaluated in the dose range 2–10 Gy (maximum uncertainty of 4.9% for 2 Gy, while minimum uncertainty of 1.4% for 10 Gy), which implies that the alanine/ESR dosimetry system can be applied to radiotherapy dose level that needs a global accuracy of 5%.