Reference ranges and biological variation of free and total serum indoxyl- and p-cresyl sulphate measured with a rapid UPLC fluorescence detection method
چکیده انگلیسی

IntroductionThe uremic toxins indoxyl sulphate (IS) and p-cresyl sulphate (pCS) are absorbed bacterial metabolites of tryptophan and tyrosine respectively and may be predictive of clinical outcome. Long chromatography times, incomplete data on the reference ranges of the free and total fractions and the biological variation limit wider clinical application.MethodsAn UPLC method with fluorescence detection was developed and reference ranges and biological variation were investigated in healthy volunteers.ResultsChromatography time was 3 min with excellent linearity, precision and low detection limits (IS of 0.02 μmol/L and pCS of 0.05 μmol/L). Both IS and pCS increased with a decrease in renal function and were moderately correlated with eGFR (R2 0.65 and 0.33 respectively). The serum reference ranges were (μmol/L): total IS of 0.7–6.3; free IS of 0.0–0.2; total pCS of 0.0–38.4; and free pCS of 0.1–2.4. The intra individual biological variation was estimated at 35.9% and 50.5% with a critical difference of 3.9 μmol/L (100%) and 20.7 μmol/L (141%) for total IS and pCS respectively.ConclusionWe describe a robust analytical method with a short chromatography time that quantifies both IS and pCS. The data on reference ranges and intra-individual biological variation need to be considered in clinical studies that investigate these uremic toxins.


► An UPLC method was developed for indoxyl sulphate (IS) and p-cresyl sulphate (pCS).
► Chromatography time was 3 min with good linearity, precision and detection limits.
► IS reference ranges (μmol/L): total of 0.7–6.3; and free of 0.0–0.2
► pCS reference ranges (μmol/L): total of 0.0–38.4; and free pCS of 0.1–2.4
► The critical difference was 100% for total IS and 150% for total pCS.