BackgroundUsing non-esterified cholesterol standard, Brevibacterium and Streptomyces are found as suitable sources of cholesterol oxidase for kinetic cholesterol assay. For clinical use, we investigated the suitability of these enzymes for cholesterol determination in human serum.MethodsWe compared the performance of reagents containing 2 enzymes for the kinetic determination of total serum cholesterol with the standardized endpoint method.ResultsReagent containing Streptomyces enzyme was more sensitive than that of Brevibacterium, with linearity up to 20.7 and 2.6 mmol/l, respectively. The analytical reaction for Streptomyces showed a shorter lag phase (148 s) and a steeper slope (absorbance vs. time) than that of Brevibacterium (246 s). The assay using Streptomyces reagent was precise and accurate and compared favorably with the endpoint method (y = 1.06x − 0.15, r = 0.996, bias = 0.21 mmol/l). Hemoglobin as high as 7.5 g/l did not interfere while turbidity greater than 2+ (absorbance > 0.778 at 670 nm) and bilirubin concentrations > 171.0 μmol/l did interfere (in a negative interference). Reagent was stable up to at least 8 weeks.ConclusionsThe Streptomyces cholesterol oxidase, with 3,4-dichlorophenol, proved a suitable source for serum total cholesterol determination by the kinetic method.