Evaluation of Hemo Techt NS-Plus system for use in a province-wide colorectal cancer screening program
موضوعات مرتبط
علوم زیستی و بیوفناوری بیوشیمی، ژنتیک و زیست شناسی مولکولی زیست شیمی
پیش نمایش صفحه اول مقاله
Evaluation of Hemo Techt NS-Plus system for use in a province-wide colorectal cancer screening program
چکیده انگلیسی

ObjectivesThe NS-Plus automated analyzer and fecal immunochemical testing (FIT) testing system (Alfresa Pharma) was evaluated for use in Newfoundland and Labrador's provincial colorectal cancer (CRC) screening program.Design and methodsVarious method performance characteristics were evaluated including the sample stability. The sensitivity for detecting neoplastic lesions was evaluated in 249 patients scheduled for colonoscopy. Each patient collected up to 2 samples for both guaiac based testing (Hemoccult ® SENSA; gFOBT) and FIT using the NS-plus system (cutoff = 20 μg Hb/g feces or 100 μg Hb/L) over 2 days. Data was analyzed comparing 1- and 2-day testing strategies.ResultsThe analyzer showed acceptable linearity, precision, and accuracy. The collection device maintained acceptable sample stability for at least 7 days at: 37 °C, room temperature (~ 23 °C), 4–8 °C, and − 20 °C. The 2-day sampling strategy identified 30% (21 of 69) of all neoplastic lesions (low and high grade adenomas and CRC) including 2 of 4 high-grade adenomas and 2 of 2 CRCs. The single day strategy identified the same high-grade adenomas and CRCs but fewer low-grade adenomas (23% of all neoplasia). Reducing the screening cutoff to the estimated 95th percentile of FIT results in the healthy adult population (10 μg Hb/g feces), detected all high-grade adenomas in the 2-day strategy.ConclusionsThe NS Plus automated analyzer system detects clinically significant neoplasms and shows acceptable performance for use in a CRC screening program with the potential for gains in sensitivity by modifying the number of days of screening or through lowering the cutoff.


► We examine NS Plus automated analyzer system for cancer screening program.
► Collection vials maintain acceptable sample stability for at least 7 days.
► Lowering cutoff or increasing screening days increases sensitivity for neoplasm.
► The NS Plus automated analyzer system detects clinically significant neoplasm.