ObjectivesLp-PLA2 is a biomarker with promise for predicting cardiac risk. The lack of reproducible results has limited its use. In evaluating a new reagent kit, we investigated conditions for optimal reproducibility.MethodsThe Auto-PLAC reagents were evaluated on the Cobas instrument. Performance characteristics, stability, and population ranges were determined.ResultsAnalytical performance characteristics replicated manufacturer's claims. The stability profile of the analyte was unusual, with increasing results observed with storage at 4 °C or − 20 °C. Only storage at − 70 °C gave acceptable stability. Population median values with properly preserved samples were much lower than the cut off previously validated for increased risk.ConclusionsIt is postulated that variability in specimen handling was a major contributor to the lack of traceability of the current reagents to the earlier clinical studies demonstrating its utility. We are now unsure how to identify reliable criteria for result interpretation.
Journal: Clinical Biochemistry - Volume 44, Issues 14–15, October 2011, Pages 1247–1252