کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1995480 | 1064974 | 2006 | 4 صفحه PDF | دانلود رایگان |

The aim of this study was to establish the degree to which a standardized test based on laser Doppler blood flow measurement is dependent on the particular equipment set-up being used. For this purpose, we examined finger skin blood flow with laser Doppler instruments in 20 healthy subjects. In laser Doppler perfusion monitoring (LDPM), we used a custom-made probe with two detecting fibers placed 0.25 and 1.2 mm from the illuminating fiber, respectively, and two laser Doppler perfusion imagers (LDPI) with a wavelength of 632.8 nm and 780 nm, respectively. Warming of the hand was achieved with a Peltier element, and reflex vasoconstriction was induced by immersing the other hand for 3 min into a water bath kept at 15°C. As a measure for the change in skin blood flow, a vasoconstriction index (VAC: cooling/before cooling) was calculated and used for the comparison of the different devices. VAC values gathered around 0.6 for all devices. However, LDPI with a wavelength of 632.8 nm showed a slightly higher VAC index, and the difference was significant. We conclude that using a standardized test is the most appropriate for monitoring changes in blood flow rather than recording and comparing discrete values in intermittent recordings. Although a difference was noted when comparing the devices, different fiber separations and wavelengths seem then to be of little consequence.
Journal: Microvascular Research - Volume 71, Issue 1, January 2006, Pages 64–67