کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
218975 463231 2013 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Voltammetric determination of verapamil and propranolol using a glassy carbon electrode modified with functionalized multiwalled carbon nanotubes within a poly (allylamine hydrochloride) film
موضوعات مرتبط
مهندسی و علوم پایه مهندسی شیمی مهندسی شیمی (عمومی)
پیش نمایش صفحه اول مقاله
Voltammetric determination of verapamil and propranolol using a glassy carbon electrode modified with functionalized multiwalled carbon nanotubes within a poly (allylamine hydrochloride) film
چکیده انگلیسی


• A GCE modified with MWCNT within a PAH was used to develop the novel sensor.
• VPM is determined by square-wave adsorptive anodic stripping voltammetry.
• PROP is determined by differential-pulse adsorptive anodic stripping voltammetry.
• The MWCNTs presented an electrocatalytic activity and increased analytical signal.
• The obtained results are in agreement with those obtained using a comparative methods.

In this report, we describe the individual determination of verapamil (VPM) and propranolol (PROP) in pharmaceutical formulations using square-wave adsorptive anodic stripping voltammetry (SWAdASV), differential-pulse adsorptive anodic stripping voltammetry (DPAdASV), and a glassy carbon electrode modified with functionalized multiwalled carbon nanotubes within a poly (allylamine hydrochloride) film. The anodic peak potential for VPM and PROP at this modified electrode was 1.09 V and 0.930 V (versus Ag/AgCl (3.0 mol L−1 KCl)), respectively, in a 0.25 mol L−1 H2SO4 solution. Under optimized SWAdASV and DPAdASV conditions, the obtained analytical curves were linear for VPM concentrations ranging from 5.0 × 10−9 to 1.1 × 10−7 mol L−1, and PROP concentrations ranging from 7.4 × 10−8 to 1.0 × 10−6 mol L−1, with detection limits of 3.6 × 10−9 mol L−1 and 2.6 × 10−8 mol L−1, respectively. Both of the proposed methods were successfully applied for the individual determination of VPM and PROP in several pharmaceutical formulations. The results were in a close agreement with a 95% confidence level for those obtained using official spectrophotometric methods.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Electroanalytical Chemistry - Volume 708, 1 November 2013, Pages 73–79
نویسندگان
, , , , ,