BackgroundEsophageal injury following catheter ablation of atrial fibrillation (AF) is reported to occur in 35% of patients. Even with a low energy setting (20–25 W), lesions develop in 10% of patients. Body mass index (BMI) has been reported to be a predictor of esophageal injury, indicating that patients with a low BMI (<24.9 kg/m2) are at a higher risk. We hypothesized that catheter ablation with a lower energy setting of 20 W controlled by esophageal temperature monitoring (ETM) at 39 °C could prevent esophageal injury even in patients with a BMI <24.9 kg/m2.MethodsTwenty patients with AF were included (age, 63±8 years; BMI, 22.9±1.3 kg/m2, left atrium diameter, 44±11 mm). If the esophageal temperature probe registered a temperature of >39 °C, radiofrequency (RF) application was stopped immediately. RF application could be performed in a “point by point” manner for a maximum of 20 s. Endoscopy was performed 1–5 days after ablation.ResultsEsophageal mucosal injury was not observed in any patient in the study.ConclusionsCatheter ablation using ETM reduced the incidence of esophageal injuries, even in patients with a low BMI.
Journal: Journal of Arrhythmia - Volume 31, Issue 1, February 2015, Pages 12–17