BackgroundCardiac CT angiography (CCTA) has become a frequently used diagnostic tool in clinical practice, but concern remains about the radiation exposure. Because of the second x-ray acquisition system, dual-source CT systems might allow for high-pitch CT data acquisition and thus for examination of the whole heart during a single heart beat, with the potential for radiation dose reduction.ObjectiveWe assessed the feasibility of a high-pitch scan mode with a dual-source CT system.MethodsHigh-pitch modes were used in patients undergoing CCTA with a dual-source CT system. Diagnostic image quality for cardiac structures and coronary arteries was assessed. Radiation dose was estimated from the scanner-generated dose-length product (DLP).ResultsCCTA was performed in 14 patients during a single heart beat applying a pitch value of 3.4. Mean heart rate during examination was 56.4 ± 8.1 beats/min. Diagnostic image quality for the assessment of larger cardiac structures was obtained in all patients, whereas diagnostic image quality could be achieved in 82% of all coronary segments. With a mean DLP of 145 ± 47 mGy × cm, the resulting estimated radiation dose was 2.0 ± 0.7 mSv.ConclusionsThis proof-of-concept study shows the ability of dual-source CT scanners to scan the whole heart during one single heart beat at low radiation dose.
Journal: Journal of Cardiovascular Computed Tomography - Volume 3, Issue 4, July–August 2009, Pages 236–242