IntroductionPressure monitoring is crucial for effective pressure therapy. Precise and reliable interface pressure measurement system, however, remains unavailable in current practice.MethodologyThis is a validation study on the application of a recently developed device, the Pliance X System, for static interface pressure measurement. Sensor properties were evaluated through series of laboratory tests in the first phase of the study. Phase II was a clinical study using the Pliance X system to differentiate the loading generated on the patients’ scars through additional inserts and pressure garment.ResultsResults showed high test–retest and inter-rater reliability (ICC ≥ 0.995) with good linearity (adjusted R-square = 0.997) by measuring standard weights. The maximum deviation between the measurement generated by the sensor and the sphygmomanometer cuff was ±1.451 mmHg. Results in the clinical trial revealed its discriminant ability in distinguishing different levels of pressure loading on patients with scars (p < 0.01).ConclusionThe commercially available pressure measurement system was found to be a reliable tool for measurement of low interface pressure under static condition.
Journal: Burns - Volume 35, Issue 6, September 2009, Pages 845–851