کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
3167529 1199366 2010 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
An ex vivo comparison of root canal length determination by three electronic apex locators at positions short of the apical foramen
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی دندانپزشکی، جراحی دهان و پزشکی
پیش نمایش صفحه اول مقاله
An ex vivo comparison of root canal length determination by three electronic apex locators at positions short of the apical foramen
چکیده انگلیسی

ObjectiveThe aim of this study was to evaluate the precision of working length determination of 3 electronic apex locators (EALs): Root ZX, RomiApex D-30, and Ipex at 0.0 mm, at the apical foramen (AF), and at 1.0 mm short of the AF.MethodologyThirty-eight mandibular premolars had their real lengths previously determined. Electronic measurements were determined at 1.0 mm, followed by measurements at 0.0 mm, performed in triplicate.ResultsPrecision of devices at 1.0 mm and 0.0 mm were: 94.7% and 97.4%, respectively (Root ZX); 78.9% and 97.4% (RomiApex D-30); and 76.3% and 97.4% (Ipex). Although no statistical differences were observed between the EALs at 0.0, at 1.0 mm Root ZX performed significantly better than the others.ConclusionThe EALs had acceptable precision when measuring the working length at the AF. However, when used at levels short of the AF, only Root ZX did not suffer a significant negative effect on precision.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology - Volume 110, Issue 2, August 2010, Pages e57–e61
نویسندگان
, , , , , , ,