Analysis of photostimulable phosphor plate image artifacts in an oral and maxillofacial radiology department
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی دندانپزشکی، جراحی دهان و پزشکی
پیش نمایش صفحه اول مقاله
Analysis of photostimulable phosphor plate image artifacts in an oral and maxillofacial radiology department
چکیده انگلیسی

ObjectiveTo analyze image artifact types and occurrence frequency when using a phosphor storage plate (PSP) digital radiographic system.Study designA total of 15,912 scanned digital images were evaluated by 3 observers, and image artifacts were classified into: 1) operator errors; 2) scanning errors; 3) PSP plate defects. To avoid damage to the sensor plate, a modification technique innovated by Roberts and Mol was used where needed and 2 double-sided pieces of tape were placed around the sensor plate covering.ResultsA total of 643 image artifacts were identified. The main image artifact cause was operator error (n = 554), followed by defects of plate (n = 60) and scanning (n = 29). Scanning errors could generally be corrected by rescanning, but most other artifacts required image retaking. The use of a modified clinical technique greatly reduced artifacts caused by wearing of the sensor plate (20%).ConclusionsModified methods to decrease the occurrence of image artifacts using PSP digital radiographic system are encouraged.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology - Volume 106, Issue 5, November 2008, Pages 749–756