کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
3169209 1199461 2007 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Validity of a newly developed ultraminiature cordless EMG measurement system
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی دندانپزشکی، جراحی دهان و پزشکی
پیش نمایش صفحه اول مقاله
Validity of a newly developed ultraminiature cordless EMG measurement system
چکیده انگلیسی

To verify validity of a newly developed ultraminiature EMG measurement system (BMS), the ability of BMS to record masseteric EMG was compared with that of a conventional polygraph system (PG) in the daytime. Effective distance between the transmitter unit and receiver unit of BMS was also examined. Subjects were 12 healthy volunteers. During tapping, maximum clenching, and gum chewing of all subjects, distinct bursts were observed in EMG recorded by BMS as well as PG. RMS values of maximum clenching measured by BMS and PG showed a linear and significant correlation, and there was no significant difference between the data of BMS and PG. When distance between the transmitter unit and receiver unit of BMS was 100 cm or less, no artifact signal was observed. Having obtained these findings suggesting ability for precise measurement in the daytime, we are planning to use BMS in home sleep studies in the next step.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology - Volume 104, Issue 5, November 2007, Pages e22–e27
نویسندگان
, , ,