کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
3169314 | 1199463 | 2007 | 4 صفحه PDF | دانلود رایگان |

ObjectiveThe aim of this study was to perform a comparative analysis of precision of 3 apex locators: Root ZX, Elements Diagnostic Unit and Apex Locator, and RomiAPEX D-30.Study designForty extracted single-rooted human teeth were selected. After endodontic access, measurement of the anatomical tooth length was visually performed by insertion of a K-file size 10 until its tip could be observed at the apical foramen with aid of a light microscope at 8× magnification. Following, the teeth were placed in a plastic box with alginate and electronically measured with the 3 apex locators at 1-mm short of the apical foramen.ResultsThe results revealed a precision rate of 97.5% for Root ZX, 95% for Elements Diagnostic Unit and Apex Locator, and 92.5% for RomiAPEX D-30, with no statistically significant difference between them (chi-square test, P > .05).ConclusionThe results confirm that all these electronic devices can accurately determine the root canal length within 1 mm from the apical constriction.
Journal: Oral Surgery, Oral Medicine, Oral Pathology, Oral Radiology, and Endodontology - Volume 104, Issue 4, October 2007, Pages e91–e94