کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
3242767 1206127 2006 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Biomechanical analysis of the percutaneous compression plate and sliding hip screw in intracapsular hip fractures: Experimental assessment using synthetic and cadaver bones
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی طب اورژانس
پیش نمایش صفحه اول مقاله
Biomechanical analysis of the percutaneous compression plate and sliding hip screw in intracapsular hip fractures: Experimental assessment using synthetic and cadaver bones
چکیده انگلیسی

SummaryWe compared the mechanical behaviour of osteosynthesis with the percutaneous compression plate (PCCP) compared with the standard osteosynthesis sliding hip screw (SHS) in intracapsular hip fractures. We created 10 stable and 10 unstable intracapsular hip fractures in 20 synthetic femurs. Each fracture was fixed with either the SHS or PCCP. In six pairs of cadaver femurs, we created unstable intracapsular hip fractures and fixed them with the SHS or PCCP, at random on the left or right side. All femoral heads were exposed to a cyclic, combined axial and torque load until failure. In each group, the PCCP resisted a significantly higher load than the SHS. Clinical prospective studies are needed to confirm these in vitro findings that the PCCP is more stable than the SHS.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Injury - Volume 37, Issue 10, October 2006, Pages 979–983
نویسندگان
, , , ,