کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
3370136 | 1219067 | 2006 | 4 صفحه PDF | دانلود رایگان |

BackgroundINNO-LiPA HBV Genotyping (LiPA HBV GT) and INNO-LiPA HBV PreCore (LiPA HBV PC) are commercially available assays for hepatitis B virus (HBV) characterization. These assays are labor-intensive and may be prone to exogenous DNA contamination due to their use of nested PCR amplification procedures and lack of contamination control measures.ObjectiveStandardized, single-round INNO-LiPA PCR amplification protocols incorporating uracil N-glycosylase and automated sample processing by the MagNA Pure LC instrument were evaluated.Study designHBV standards containing 10,000, 1000, 100, 10, and 0 IU/mL were analyzed to determine the analytical sensitivity and reproducibility of these modified procedures. One hundred clinical serum specimens with viral titers ranging from 390 to 16,900,000 IU/mL were tested by modified LiPA HBV GT, while 34 specimens with viral titers ranging from 378 to 11,600,000 IU/mL were tested by modified LiPA HBV PC.ResultsModified LiPA HBV GT and LiPA HBV PC each yielded analytical sensitivities of 100% at an HBV DNA level of 1000 IU/mL and 90% at a level of 100 IU/mL. Among clinical specimens, success rates for both INNO-LiPA procedures were ≥94%.ConclusionsBoth modified INNO-LiPA procedures were sensitive and reproducible, with improved efficiency and suitability for routine laboratory use.
Journal: Journal of Clinical Virology - Volume 37, Issue 3, November 2006, Pages 218–221