The outcome of a single design of dual mobility cup was prospectively evaluated in a continuous series of 994 revision THAs with respect to dislocation and intra-prosthetic dislocation (IPD). At a 7.3-year mean follow-up, the dislocation rate was 1.5% and the IPD rate was 0.2%. The 2 IPD occurred in acetabular-only revisions and were related to a poor head-to-neck ratio with early impingement and wear at the polyethylene mobile component chamfer. Dual mobility cups demonstrated a low dislocation rate in revision THA but did not compensate for potential perioperative technical errors. In addition, IPD did not appear to be a concern with respect to the benefit in term of instability prevention though caution is advised in acetabular-only revision associated with a poor head-to-neck ratio. Level of Evidence: Therapeutic study—Level IV