کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
4060686 | 1603999 | 2014 | 5 صفحه PDF | دانلود رایگان |

We report the results of 58 hip resurfacing arthroplasties (HRA) revised by a single surgeon with an average of 5.2 ± 2.6 years follow-up. The four most common causes for revision were acetabular component loosening, femoral neck fracture, femoral component loosening, and adverse wear related failure (AWRF). In 95% of cases (55/58), the revision bearing was a large metal-on-metal type including all seven AWRF cases; three cases were revised to ceramic-on-polyethylene. There were two repeat revisions due to acetabular component loosening. Revision of AWRF had an excellent outcome using limited debridement and a stable large metal bearing placed in the correct position. The only problematic group was the one revised for acetabular loosening in which 2/16 (6%) required repeat revision for failure of acetabular fixation.