کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4062392 1604030 2012 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Cementless Metasul Metal-On-Metal Total Hip Arthroplasties at 13 Years
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Cementless Metasul Metal-On-Metal Total Hip Arthroplasties at 13 Years
چکیده انگلیسی

Second-generation metal-on-metal bearings have been used since the late 1980s as alternative bearings to eliminate aseptic loosening due to polyethylene wear. The aim of the present study was to evaluate the long-term results of a series of Metasul (Zimmer GmbH, Winterthur, Switzerland) metal-on-metal total hip arthroplasty (THA). One hundred forty-nine cementless THAs with a 28-mm Metasul articulation were performed in 111 consecutive patients. The results were retrospectively reviewed at 13 years postoperatively. Clinical and radiographic evaluations and implant survivorship were performed. Seven hips (4.7%) were revised. The overall survivorship with revision for any reason as the end point was 0.94. The average Harris hip score was 91.4. Expansive osteolysis was found adjacent to the stem in 4 hips. Metal wear–related aseptic loosening was not the major reason for failure in our Metasul metal-on-metal THAs. Recurrent dislocation was the main reason for revision in our series.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 27, Issue 2, February 2012, Pages 186–192