چکیده انگلیسی
Modularity in sleeved femoral components allows the exchange of the stem without disruption of the fixation between the sleeve and the surrounding bone at revision surgery. Failure to disengage the stem from the sleeve would represent an unnecessary compromise from the intended usefulness of the modular design. We report the results of an examination of 22 modular titanium alloy femoral components retrieved after 0.0 to 8.8 years in vivo. In 7 implants, the stem-sleeve interface could not be disengaged without cutting through the components or using mechanical force. Moderate to severe corrosion was detected in all 7 of these cases. Corrosive surface changes were observed in an additional 6 interfaces. There was no correlation with the length of time that the devices had been implanted. When only the stem is to be revised, orthopedic surgeons should be aware of difficulties in disengagement and anticipate alternative surgical procedures.