کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4063182 1604037 2011 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Second-Generation Locking Mechanisms and Ethylene Oxide Sterilization Reduce Tibial Insert Backside Damage in Total Knee Arthroplasty
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Second-Generation Locking Mechanisms and Ethylene Oxide Sterilization Reduce Tibial Insert Backside Damage in Total Knee Arthroplasty
چکیده انگلیسی

This study evaluated the effects of polyethylene quality and locking mechanism on damage to the nonarticulating (backside) surface of retrieved tibial inserts in total knee arthroplasty. Inserts with peripheral capture (PC) locking mechanisms and ethylene oxide (EtO)-sterilized polyethylene were hypothesized to prevent major backside damage. A total of 156 inserts were sorted by locking mechanism and sterilization method and analyzed by damage scoring methods. Ninety-seven specimens exhibited burnishing. Significant positive linear correlations were observed between damage score and age in vivo for all combinations, but damage occurred at a significantly lower rate for second-generation PC implants with EtO sterilization. Most specimens in this group were undamaged (46/72), with others exhibiting only burnishing. Sex, body mass index, and weight did not influence backside damage.