چکیده انگلیسی

Press-fit acetabular cups without screw holes can limit migration of particulate wear debris and reduce risk of acetabular osteolysis and device loosening. The Tri-Spike cup (Biomet, Inc, Warsaw, Ind) includes a titanium alloy plasma spray porous surface and does not require screw fixation. We retrospectively examined the incidence of cup loosening and acetabular osteolysis after implantation of 45 cups (44 patients) with mean follow-up of 7.3 years (range, 4-9 years). Only one patient (one cup) had evidence of less than 1 mm of retroacetabular radiolucency at 3 years (nonprogressive), which was found to remain firmly fixed during revision of the aseptically loosened femoral component. No cups were removed or revised at latest follow-up. Projected Kaplan-Meier survivorship at 9 years was 100% for cup loosening/revision and 97.8% for radiolucency.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 24, Issue 3, April 2009, Pages 391–399