کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4063329 1604039 2011 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Metal Ion Release With Large-Diameter Metal-on-Metal Hip Arthroplasty
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
پیش نمایش صفحه اول مقاله
Metal Ion Release With Large-Diameter Metal-on-Metal Hip Arthroplasty
چکیده انگلیسی

Preoperative and postoperative ion concentrations were measured in 29 metal-on-metal, large-diameter head total hip arthroplasty (LDH-THA) patients. Mean chromium, cobalt (Co), and titanium levels from LDH-THA were 1.3, 2.2, and 2.7 μg/L at 12 months. The open femoral head design showed significantly higher Co concentrations than the closed design (3.0 vs 1.8 μg/L, P = .037). Compared with previously published ion levels from a hip resurfacing system presenting the same bearing characteristics, Co levels were significantly higher in LDH-THA (2.2 vs 0.7 μg/L, P < .001). This study has demonstrated that the addition of a sleeve with modular junctions and an open femoral head design of LDH-THA causes more Co release than bearing surface wear (157% and 67%, respectively). Even if no pathologic metal ion threshold level has been determined, efforts should be made to minimize its release. We recommend modification or abandonment of the modular junction and femoral head open design for this specific LDH-THA system.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: The Journal of Arthroplasty - Volume 26, Issue 2, February 2011, Pages 282–288