Reflectance spectrometry for real-time hemoglobin determination of placental vessels during endoscopic laser surgery for twin-to-twin transfusion syndrome
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی پریناتولوژی (پزشکی مادر و جنین)، طب اطفال و بهداشت کودک
پیش نمایش صفحه اول مقاله
Reflectance spectrometry for real-time hemoglobin determination of placental vessels during endoscopic laser surgery for twin-to-twin transfusion syndrome
چکیده انگلیسی

ObjectiveThe objective of the study was to develop a noninvasive technique to determine hemoglobin (Hb) content through spectral analysis of diffusely reflected broadband visible illumination from individual blood vessels during endoscopic fetal surgery for twin-to-twin transfusion syndrome (TTTS).MethodsThe reflection of an incoming xenon endoscopic light source was captured through a 630-μm–diameter optic fiber coupled to a fixed grating spectrometer (2-nm resolution). A 450- to 700-nm wavelength range was used for analysis. Three data-capturing methods were studied: (1) fixed-image spectrum capture with fiber aimed at (but not touching) center of a vessel, (2) no-touch scanning perpendicular to the vessel and dynamic spectral capture, and (3) dynamic spectral capture and analysis of the reflectance spectra during brief vessel touch.ResultsEight controls (elective laparoscopic and thoracoscopic operations in children aged 1-17 years) were enrolled. Four vessels were analyzed in each case. The brief-touch technique with intensity peak analysis yielded the most reproducible results between multiple vessels in the same patient. Spectrometry was also applied to 2 TTTS patients. The (anemic) donor and (polycythemic) recipient twin fetuses could be differentiated with good correlation between vessels (arteries and vein) of the same fetus.ConclusionsIt is possible to differentiate donor from recipient placental vessels by spectral analysis of the reflected light through the endoscope using a noninvasive and real-time method. This may improve the accuracy of endoscopic laser ablation of placental vessels in TTTS and may allow instant endoscopic Hb determination for laparoscopic procedures as well.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Pediatric Surgery - Volume 45, Issue 1, January 2010, Pages 59–64
نویسندگان
, , , , , , ,