کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4430151 1619849 2011 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Validation of analysis of arsenic in water samples using Wagtech Digital Arsenator
موضوعات مرتبط
علوم زیستی و بیوفناوری علوم محیط زیست شیمی زیست محیطی
پیش نمایش صفحه اول مقاله
Validation of analysis of arsenic in water samples using Wagtech Digital Arsenator
چکیده انگلیسی

Groundwater, the only source of potable water for millions of people in Bangladesh during dry season, is often contaminated with arsenic (As) above the allowable drinking water limit of 50 μg/L. Testing well water – with arsenic field test kits (FTKs) – and switching to safe-wells can effectively reduce exposure to As. FTKs are low cost, provide fast results, and are commercially available. There are between 10 and 11 million shallow tubewells in Bangladesh. Approximately, 5 million have been tested using FTKs. FTKs with color comparator rely on visual identification for generating qualitative results, which may not be highly reliable at the lower range because human eyes have low sensitivity in this range and sensitivity varies from person to person. The Wagtech Digital Arsenator does not suffer from this limitation and should, in theory, be able to generate quantitative, accurate, and reliable results. The instrument has a linear range of 0–100 μg/L, an accuracy of ± 20% for the 50 μg/L quality control standards, and a detection limit of about 4.4 μg/L. All Arsenators employed in this investigation also displayed high bias for 50 μg/L arsenic standard and were calibrated in order to improve measurement accuracy and reliability. Analyses of 179 raw and 92 treated well waters in the field and in two analytical laboratories were found to be highly correlated with the Spearman rank correlation coefficient of ≥ 0.800, indicating that Arsenator results are perhaps nearly as accurate and reliable as those from analytical laboratories.

Research Highlights
► All Arsenators displayed high bias and need calibration for accurate and reliable results.
► Arsenator has a linear range of 0 – 100 μg/L, accuracy of ± 20, and detection limit of 4.4 μg/L.
► The instrument results and accuracy is comparable to those from analytical laboratories.
► Arsenator would be a valuable test kit for well testing and identification of safe wells.