کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4452981 1312121 2009 12 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Design and performance of an optimized electrical diffusion battery
موضوعات مرتبط
مهندسی و علوم پایه علوم زمین و سیارات علم هواشناسی
پیش نمایش صفحه اول مقاله
Design and performance of an optimized electrical diffusion battery
چکیده انگلیسی

Standard aerosol instruments to measure particle size distributions in the ultrafine size range are large and heavy. We built a small electrical diffusion battery, which can be carried in a backpack and thus makes size-resolved short-term personal aerosol exposition measurements possible. The instrument was designed for maximal measurable particle size range and long maintenance intervals. The optimum number of stages for the diffusion battery was found with a Monte-Carlo simulation. To assess the instrument's performance, we compared size distributions measured with the electrical diffusion battery to those obtained with standard aerosol instruments (the scanning mobility particle sizer and the fast mobility particle sizer of TSI Inc.). In general, the readings of the electrical diffusion battery and those of standard instruments agree well (to within 10–20%).

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Aerosol Science - Volume 40, Issue 2, February 2009, Pages 152–163
نویسندگان
, , ,