Cluster ions have been recognized as a superb primary species in time of flight secondary ion mass spectroscopy (ToF-SIMS) compared with monatomic primary ions, as they significantly enhance the secondary ion yields from bulk samples. Self-assembled monolayers provide an important system for studying the fundamental mechanism involved in the yield enhancement.We used a gold cluster ion source to analyze a new type of self-assembled monolayer: a fluorocarbon-grafted polyethylene terephthalate. In addition to the structure details, which helped to understand the grafting mechanism, ToF-SIMS analysis revealed that fluorocarbon secondary ion yield enhancements by cluster ions were due to the enhanced sputter efficiency. A larger information depth may also be expected from the enhancement. Both mathematical definitions of damage cross-section and disappearance cross-section were revisited under a new context. Another cross-section parameter, sputter cross-section, was introduced to differentiate the beam induced sputter process from damage process.