کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
5509602 1538627 2017 30 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
POCT PT INR - Is it adequate for patient care? A comparison of the Roche Coaguchek XS vs. Stago Star vs. Siemens BCS in patients routinely seen in an anticoagulation clinic
موضوعات مرتبط
علوم زیستی و بیوفناوری بیوشیمی، ژنتیک و زیست شناسی مولکولی زیست شیمی
پیش نمایش صفحه اول مقاله
POCT PT INR - Is it adequate for patient care? A comparison of the Roche Coaguchek XS vs. Stago Star vs. Siemens BCS in patients routinely seen in an anticoagulation clinic
چکیده انگلیسی
The CoaguChek XS and on-site Stago analyzers which used human recombinant (ISI = 1.01) and rabbit brain thromboplastin (ISI = 1.25), respectively, showed reasonable agreement for INR < 3.0 (k = 0.62) but significant difference for INR ≥ 3.0 (k = 0.10). In contrast, the CoaguChek XS and Siemens BCS XP, which both employed human recombinant thromboplastin (BCS ISI = 1.02), showed greater agreement for the complete range INR values (INR < 3.0 k = 0.84; INR ≥ 3.0 k = 0.70). ECAA Poller calibrant data showed the automated instruments were performing as expected, indicating that ISI calibrations were correct but insufficient to standardize the INR values for the different thromboplastin methods across the full range of measured INRs. Central lab verification of POCT INR > 5.0 with the Stago Evolution prevented adverse treatment events for the warfarin therapy patients in the six months preceding and following this investigation.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Clinica Chimica Acta - Volume 472, September 2017, Pages 139-145
نویسندگان
, , , , , , ,