Ultra-sensitive conductometric detection of pesticides based on inhibition of esterase activity in Arthrospira platensis
چکیده انگلیسی
Enzymatic conductometric biosensor, using immobilized Arthrospira platensis cells on gold interdigitated electrodes, for the detection of pesticides in water, was elaborated. Cholinesterase activity (AChE) was inhibited by pesticides and a variation of the local conductivity was measured after addition of the substrate acetylthiocholine chloride (AChCl). The Michaelis-Menten constant (Km) was evaluated to be 1.8 mM through a calibration curve of AChCl. Inhibition of AChE was observed with paraoxon-methyl, parathion-methyl, triazine and diuron with a detection limit of 10−18 M, 10−20 M, 10−20 M and 10−12 M, respectively and the half maximal inhibitory concentration (IC50) was determined at 10−16 M, 10−20 M, 10−18 M and 10−06 M, respectively. An important decrease of response time τ90% was recorded for AChE response towards AChCl after 30 min cell exposure to pesticides. Scanning electron microscopy images revealed a degradation of the cell surface in presence of pesticides at 10−06 M.