An integrated microfluidic system for measurement of glycated hemoglobin Levels by using an aptamer-antibody assay on magnetic beads
ترجمه فارسی عنوان
یک سیستم میکروفلوئیدیک یکپارچه برای اندازه گیری سطح هموگلوبین گلیکوز شده با استفاده از آزمون آنتی بادی آپتامر بر روی دانه های مغناطیسی
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
چکیده انگلیسی
Blood glycated hemoglobin (HbA1c), reflecting the average blood glucose level in the proceeding 2-3 months, is recommended for screening/diagnosing and patient management of diabetes. However, accurate measurement of the HbA1c level at the point of care is hampered by costly, large-scale instruments (such as high-performance liquid chromatography) or reagent instability of classical immunologic methods, which involve antibody-based immunoturbidimetry. In this work, an integrated microfluidic system using aptamer-based testing to measure HbA1c in blood samples is therefore presented. This measuring system used nucleic-acid aptamers that exhibited high sensitivity and high specificity for hemoglobin and HbA1c to perform a stable and robust testing. The compact microfluidic system consumed less samples and reagents and significantly shortened the detection time. Combining the advantages of microfluidics and aptamers, this integrated microsystem presents a promising tool for accurate and point-of-case HbA1c detection. To demonstrate its clinical utility, whole blood samples with clinically-relevant concentrations of HbA1c and Hb were automatically measured on the integrated microfluidic system. Experimental data showed that the developed aptamer-based microfluidic system is capable of detecting HbA1c and Hb with a good linear response. The entire process was completed within 25 min. The aptamer-antibody on-chip sandwich immunoassay may be further refined to allow diabetes screening and diagnosis at lower cost and earlier phase to minimize the risk of diabetic complications.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biosensors and Bioelectronics - Volume 68, 15 June 2015, Pages 397-403
نویسندگان
, , , , ,