کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
7556822 1491298 2018 33 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Mucin and carbon nanotube-based biosensor for detection of glucose in human plasma
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Mucin and carbon nanotube-based biosensor for detection of glucose in human plasma
چکیده انگلیسی
The novel CNT-muc composite provided a sensitivity of 0.44 ± 0.01 mA M−1 and a response time of 28 ± 2 s. These values were respectively 20% higher and 40% shorter than those obtained with a sandwich-type biosensor prepared without CNT. Additionally, CNT-muc based biosensor exhibited more than 3 orders of magnitude of linear dynamic calibration range and a detection limit of 3 μM. The short-term and long-term stabilities of the biosensors were also examined and excellent results were obtained through successive experiments performed within the first 60 days from their preparation. Finally, the storage stability was remarkable during the first 300 days.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Analytical Biochemistry - Volume 550, 1 June 2018, Pages 34-40
نویسندگان
, , , ,