کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
811295 | 1469142 | 2011 | 8 صفحه PDF | دانلود رایگان |

With the emergence of new biomaterials and elastography imaging techniques, there is a need for innovative instruments dedicated to viscoelasticity measurements. In this work, we introduce a novel hyper-frequency viscoelastic spectroscopy (HFVS) technique dedicated to characterize soft media subjected to mid-to-very-high frequency stress ranges (or, equivalently, to probe short-to-very-short relaxation times). HFVS, which has been implemented in an analytical instrument performing non-contact measurements in less than 1 s between 10 and 1000 Hz, is a suitable tool to study viscoelasticity for bio-applications. In this context, HFVS has been compared to classical oscillatory rheometry on several classes of soft materials currently encountered in tissue repair, bioengineering and elastography imaging on a frequency range between 10 and 100 Hz. After having demonstrated the good correspondence between HFVS and rheometry, this study has been completed by exploring the sensitivity of HFVS to physicochemically induced variations of viscoelasticity. HFVS opens promising perspectives in the challenging field of biomaterial science and for viscoelasticity-based quality control of materials.
Journal: Journal of the Mechanical Behavior of Biomedical Materials - Volume 4, Issue 7, October 2011, Pages 1115–1122