کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
868375 909807 2009 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A stable and selective electrochemical biosensor for the liver enzyme alanine aminotransferase (ALT)
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
A stable and selective electrochemical biosensor for the liver enzyme alanine aminotransferase (ALT)
چکیده انگلیسی

An electrochemical method to determine alanine aminotransferase (ALT) activity over its normal and elevated physiological range was developed based upon detection of l-glutamate at a glutamate oxidase-modified platinum electrode. Measurements were carried out in the presence of ALT co-substrates l-alanine and α-ketoglutarate and current response from either the oxidation of hydrogen peroxide or the re-oxidation of the mediator ferrocene carboxylic acid was employed. The enzyme electrode was tested over a 6-month period and found to retain 79% of its original activity towards ALT detection with >200 measurements performed over this time. Signals associated with interfering electroactive species (ascorbic acid and uric acid) were eliminated using background subtraction at a denatured glutamate oxidase enzyme electrode. The sensitivity of the device was found to be 0.845 nA U−1 L ALT with t90 = 180 s, linear range 10–1000 U L−1 and LOD of 3.29 U L−1 using amperometry at Eapp = 0.4 V vs. Ag/AgCl at 308 K (35 °C).

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biosensors and Bioelectronics - Volume 24, Issue 9, 15 May 2009, Pages 2926–2930
نویسندگان
, , , ,