کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
868513 909809 2011 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
A contact lens with embedded sensor for monitoring tear glucose level
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
A contact lens with embedded sensor for monitoring tear glucose level
چکیده انگلیسی

We report the design, construction, and testing of a contact lens with an integrated amperometric glucose sensor, proposing the possibility of in situ human health monitoring simply by wearing a contact lens. The glucose sensor was constructed by creating microstructures on a polymer substrate, which was subsequently shaped into a contact lens. Titania sol–gel film was applied to immobilize glucose oxidase, and Nafion® was used to decrease several potential interferences (ascorbic acid, lactate, and urea) present in the tear film. The sensor exhibits a fast response (20 s), a high sensitivity (240 μA cm−2 mM−1) and a good reproducibility after testing a number of sensors. It shows good linearity for the typical range of glucose concentrations in the tear film (0.1–0.6 mM), and acceptable accuracy in the presence of interfering agents. The sensor can attain a minimum detection of less than 0.01 mM glucose.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biosensors and Bioelectronics - Volume 26, Issue 7, 15 March 2011, Pages 3290–3296
نویسندگان
, , , , ,