چکیده انگلیسی

A novel H2O2 biosensor has been fabricated based on the direct electrochemistry and electrocatalysis of myoglobin (Mb) immobilized on silver nanoparticles doped carbon nanotubes film with hybrid sol–gel techniques. A pair of redox peaks with peak separation of 160 mV and formal potential of −0.295 V was observed at this composite film, corresponding to the direct electrochemistry of Mb. The heterogeneous rate constant was estimated to be 0.41 s−1. Under optimum conditions, the amperometric determination of H2O2 was performed with a linear range of 2.0 × 10−6–1.2 × 10−3 mol L−1 and a detection limit of 3.6 × 10−7 mol/L (S/N = 3). The Michealis–Menten constant was also estimated to be 1.62 mmol L−1. The proposed biosensor showed favorable reproducibility, stability, selectivity and accuracy, and has been used to determine H2O2 in real samples with favorable recoveries.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biosensors and Bioelectronics - Volume 24, Issue 7, 15 March 2009, Pages 2149–2154