Development of a disposable glucose biosensor using electroless-plated Au/Ni/copper low electrical resistance electrodes
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Development of a disposable glucose biosensor using electroless-plated Au/Ni/copper low electrical resistance electrodes
چکیده انگلیسی

This paper presents a glucose biosensor, which was developed using a Au/Ni/copper electrode. Until now, research regarding the low electrical resistance and uniformity of this biosensor electrode has not been conducted. Glucose oxidase (GOD) immobilized on the electrode effectively plays the role of an electron shuttle, and allows glucose to be detected at 0.055 V with a dramatically reduced resistance to easily oxidizable constituents. The Au/Ni/copper electrode has a low electrical resistance, which is less than 0.01 Ω, and it may be possible to mass produce the biosensor electrode with a uniform electrical resistance. The low electrical resistance has the advantage in that the redox peak occurs at a low applied potential. Using a low operating potential (0.055 V), the GOD/Au/Ni/copper structure creates a good sensitivity to detect glucose, and efficiently excludes interferences from common coexisting substances. The GOD/Au/Ni/copper sensor exhibits a relatively short response time (about 3 s), and a sensitivity of 0.85 μA mM−1 with a linear range of buffer to 33 mM of glucose. The sensor has excellent reproducibility with a correlation coefficient of 0.9989 (n = 100 times) and a total non-linearity error of 3.17%.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Biosensors and Bioelectronics - Volume 24, Issue 3, 15 November 2008, Pages 410–414
نویسندگان
, , , , ,