• This study investigates the hemodynamic effect of catheter insertion in coronary artery.
• The WSS and velocity in LAD were notably increased by the catheter insertion.
• The WSS on the catheter was also at high level.
• These effects may increase the risk of perioperative complications.
Intravascular stenting is one of the most commonly used treatments to restore the vascular lumen and flow conditions, while perioperative complications such as thrombosis and restenosis are still nagging for patients. As the catheter with crimped stent and folded balloon is directly advanced through coronary artery during surgery, it is destined to cause interference as well as obstructive effect on blood flow. We wonder how the hemodynamic environment would be disturbed and weather these disturbances cause susceptible factors for those complications. Therefore, a realistic three-dimensional model of left coronary artery was reconstructed and blood flow patterns were numerically simulated at seven different stages in the catheter insertion process. The results revealed that the wall shear stress (WSS) and velocity in left anterior descending (LAD) were both significantly increased after catheter inserted into LAD. Besides, the WSS on the catheter, especially at the ending of the catheter, was also at high level. Compared with the condition before catheter inserted, the endothelial cells of LAD was exposed to high-WSS condition and the risk of platelet aggregation in blood flow was increased. These influences may make coronary arteries more vulnerable for perioperative complications.
Journal: Medical Engineering & Physics - Volume 38, Issue 9, September 2016, Pages 946–951