Load-transfer analysis after insertion of cementless anatomical femoral stem using pre- and post-operative CT images based patient-specific finite element analysis
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی پزشکی
پیش نمایش صفحه اول مقاله
Load-transfer analysis after insertion of cementless anatomical femoral stem using pre- and post-operative CT images based patient-specific finite element analysis
چکیده انگلیسی

Periprosthetic bone remodeling is commonly seen after total hip arthroplasty, but the remodeling pattern differs among patients even in those implanted with the same stem. Remodeling occurs mainly because of the difference in load transmitted from the stem to the femur. In this study, we evaluated the load-transfer pattern in eight female patients implanted with an anatomical stem on an individual basis by patient-specific finite element analysis that is based on pre- and postoperative computed tomography images. Load transfer was evaluated using interface stress between the stem and bone. One of eight patients demonstrated proximal dominant load transfer, while the other patients demonstrated a distal dominant pattern. The results of our biomechanical simulations reveal the differences in load-transfer pattern after surgery among patients with the same anatomical stem.