The development of needles, needle–insertion simulators, and needle-wielding robots for use in a clinical environment depends on a thorough understanding of the mechanics of needle–tissue interaction. It stands to reason that the forces arising from this interaction are influenced by numerous factors, such as needle type, insertion speed, and tissue characteristics. However, exactly how these factors influence the force is not clear. For this reason, the influence of various factors on needle insertion-force was investigated by searching literature for experimental data. This resulted in a comprehensive overview of experimental insertion-force data available in the literature, grouped by factor for quick reference. In total, 99 papers presenting such force data were found, with typical peak forces in the order of 1–10 N. The data suggest, for example, that higher velocity tends to decrease puncture force and increase friction. Furthermore, increased needle diameter was found to increase peak forces, and conical needles were found to create higher peak forces than beveled needles. However, many questions remain open for investigation, especially those concerning the influence of tissue characteristics.
Journal: Medical Engineering & Physics - Volume 34, Issue 6, July 2012, Pages 665–680